Testomat 808 SILICA

Online Analyzer for Limit Value Detection
The Testomat® 808 SiO2 has been designed for use in the sterilisation of equipment in hospitals. It complies with the EN 285:2006 standard. The device is a limit gauge that automatically monitors the level of SiO2 in the water. It gives you the choice between 10 limits in the range of 0.3 to 1.2 ppm.
Range of Application
-
sterilization of equipment in hospitals
-
silica monitoring
Measuring range:
Silica SiO2: 0.3 – 1.2 ppm
Water Supply
-
Operating pressure:
4.4 - 58 psi (option to add a pressure reducer for pressure over 58) -
Water inlet: 1/4-in inlet fitting
-
Water outlet: Opaque hose with an internal diameter of 0.55 inch
-
Water temperature: 50 to 104 °F (a cooler can be installed for temperatures above 104°F)
Technical Data
-
Power supply: 24 / 115 / 230 VAC, 50 – 60 Hz
-
Power consumption: Max. 16 V without external load
-
Current Interface Card: Not available
-
Protection class: I
-
Degree of protection: IP 44
-
Ambient temperature: 50 - 104°F
-
Dimensions: (W x H x D): 14.3 x 12.4 x 5.4 inch
-
Weight: 9.59 lb