Titromat M1
Titration Monitor for Alkalinity
The Titromat® M1 is utilized for the automatic determination and monitoring of alkalinity (minus m-value) in water. The measuring range with the reagents TC2010A and TC2010B is 0.018 - 0.36 mmol/l with a resolution of 0.009 mmol/l and an accuracy of +/- 0.04 mmol/l.
The Features Include:
-
Simple, menu-driven operating and programming via a plain text display
-
Freely selectable hardness unit in °dH, °f, ppm CaCO3, mmol/l
-
High measuring accuracy provided by a precise piston-dosing pumps
-
Analysis initiation:
-
Automatic interval operation (adjustable Interval pause 0 to 99 min.)
-
External control
-
Dynamic start (exhaustion dependent)
-
Volume controlled (water meter)
-
-
Two independent limit values with hysteresis (1, 2 or 3 bad analyses) and adjustable switch functions
-
Monitoring of two measuring points (change-over via external solenoid valves)
-
Internal error documentation
-
Programmable service address
-
Programmable maintenance interval for a maintenance request
-
Extended operating periods due to 500 ml indicator storage bottle
-
Optional:
-
Interface (0/4-20 mA or 0/2-10 V)
-
Interface RS 232
-
SD Card Datalogger
-
Water Supply
-
Operating pressure:
0.3 to 1 bar (4.4 to 15.4 psi) or
1 to 8 bar (14.5 to 116 psi) -
Water inlet: 1/4-in inlet fitting
-
Water outlet: Opaque hose with an internal diameter of 0.47 inch
-
Water temperature: 50 to 104 °F (a cooler can be installed for temperatures above 104°F)
Technical Data
-
Power supply: 24,115 and 230–240 VAC, all 50–60Hz
-
Power consumption: Max. 30 V
-
Current Interface Card: Separately
-
Protection class: I
-
Degree of protection: IP 65
-
Ambient temperature: 50 - 104°F
-
Dimensions: (W x H x D): 18.9 x 14.9 x 11 inch
-
Weight: 21 lb