top of page

Testomat® 808

Manual Testomat® 808

Flyer Testomat® 808

Maintenance Testomat® 808

Limit value measurement for applications such as reverse osmosis plants, softwater for commercial purposes, pure water production plants or galvanization

 

Measuring range
Water hardness 0.02 – 5.0 °dH

(0.4 to 54 ppm CaCO3)

MAINS WATER SUPPLY
  • Operating pressure: Depending on product configuration: 0.3 to 1 bar / 4.4 to 14.5 psi ; 1 to 4 bar / 14.5 to 58 psi
    (a pressure reducer (special accessory) should be used from 4 to 8 bar / 58 to 116 psi range)

  • Water inlet: Opaque hose with 0.24 in. external diameter/ 0.16 in. internal diameter

  • Water outlet: Hose with 0.24 in. external diameter/ 0.16 in. internal diameter

  • Water temperature: 50 to 104 °F

TECHNICAL DATA
  • Power supply: Depending on product configuration: 24V / 115V / 230V, 50 – 60 Hz

  • Instrument protection: 230 – 240 V: T0.1 A; 115 V: T0.2 A; 24 V: T0.8 A

  • Main protection for consumers: max. 4 A (n , l)

  • Power consumption: max. 16 VA, without external load

  • Protection class: I

  • Degree of protection: IP 54

  • Conformity: EN 61000-6-2, EN 61000-6-4, EN 61010-1

  • Ambient temperature: 50 - 104°F

  • Current interface: Output of defined values (5, 8, 11, 14, 17, 20 mA) for displaying status and error messages, max. load 500 Ohms

  • Contact load relay: 230V / 4A AC ohm resistive load

  • Dimensions: W x H x D = 14.3 x 12.4 x 5.4 in.

  • Weight: 9.59 lb.

Indicators for Testomat® 808